Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/17/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Medical devices: targeting guide catalog#: 00249000300 lot#: 62186171.It is unknown if the product will be returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It is reported that during a procedure while the surgeon attempted to insert the nail the threads of the impaction head fractured in the guide.No adverse events were a result of this malfunction.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
(b)(4).Complaint sample was evaluated and the reported event was confirmed.Visual inspection indicated the threads of the impactor head have fractured off and are retained in the jig handle.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|