ETHICON ENDO-SURGERY, LLC. HARMONIC LAP 5MM SHEAR 36CM; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HAR36
Device Problems Failure to Form Staple (2579); Failure to Cut (2587)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/25/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # r9334g.Investigation summary: the device was returned with no apparent damage.The device was assembled and disassembled to a test hand piece without any difficulty.The device was tested on a gen11 and an alert screen was displayed.The instrument was disassembled to inspect the internal components and it was noticed that the blade sheath was missing.The preliminary investigation into the missing sheath issue has identified our manufacturing process as the possible cause.The batch history record was reviewed and there were no defects, protocols or ncr(s) found during the manufacturing process related to this complaint.
 
Event Description
It was reported that the initial linking was without success.Scissors did not work.Third scissor worked.No patient consequence.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC LAP 5MM SHEAR 36CM
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7965880
MDR Text Key123951978
Report Number3005075853-2018-13607
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10705036014447
UDI-Public10705036014447
Combination Product (y/n)N
PMA/PMN Number
K120729
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2023
Device Catalogue NumberHAR36
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2018
Initial Date Manufacturer Received 09/18/2018
Initial Date FDA Received10/15/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/23/2018
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
-
-