• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD ADULT NASAL INTERFACE; CAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD ADULT NASAL INTERFACE; CAT Back to Search Results
Model Number OPT944
Device Problem Device Handling Problem (3265)
Patient Problem Pressure Sores (2326)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The opt944 interface is used to deliver humidified oxygen to patients and consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Method: the complaint opt944 nasal cannula was not returned to fisher & paykel healthcare in (b)(4).Therefore, our investigation was performed based on the information provided by the customer and our knowledge of the product.Conclusion: we were unable to confirm what caused the pressure sores on the patient's ear.However, based on our knowledge of the product, pressure sores are often a result of incorrect set up.Fisher & paykel healthcare representative offered to provide further training regarding correct fitting and set up to the customer, however it was declined.The user instructions which accompany the opt944 nasal cannula contain pictorial instructions for setting up the nasal cannula.The instructions also state that the interface is intended to be used "for delivery of humidified respiratory gases" and includes the following cautions/warnings: - do not crush or stretch tube, to prevent loss of therapy.- failure to use the set-up described above can compromise performance and affect patient safety.
 
Event Description
A healthcare facility in (b)(6) reported via a fisher & paykel healthcare (f&p) representative that the optiflow nasal cannula opt944 caused grade 3 pressure sores above ears on a patient.There was no further patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT NASAL INTERFACE
Type of Device
CAT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9194534000
MDR Report Key7976598
MDR Text Key124829594
Report Number9611451-2018-00894
Device Sequence Number1
Product Code CAT
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT944
Device Catalogue NumberOPT944
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/20/2018
Initial Date FDA Received10/17/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-