ST. JUDE MEDICAL, INC. MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER Back to Search Results
Model Number EX1150
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/10/2018
Event Type  malfunction  
Event Description
After plugging the transmitter into power, the power supply was observed to emit smoke and sparks.The issue was resolved by removing and replacing the transmitter.The patient was in stable condition and experienced no adverse health consequences.
 
Manufacturer Narrative
The field complaint of the transmitter's power supply emitting smoke and sparks was not verified as the event could not be reproduced; however, there was burn damage revealed that could have been the initial cause of the smoke.The visual inspection revealed no physical damage to the outer housing of the merlin@home transmitter and no physical damage to the outer casing of the ac power adapter.The prongs were removed from the ac power adapter to inspect for damage, and a strong burnt odor was found emitting from the ac power adapter.Thus the unit was not plugged into a working ac electrical outlet.Upon opening the ac power adapter, inspection revealed that there were burnt components on both sides of the main pcb and on its¿ inner casing.Inspection of the opened transmitter revealed that there was no damage to the main pcb or to its inner housing.The original ac power adapter was replaced to test the transmitter with the ac power adapter tl# (b)(4).The unit was then plugged into a working ac electrical wall outlet and the power on function was performed successfully.The test revealed that the burnt damage to the original ac power adapter was the cause of the no power issue.As a result, we can conclude that the transmitter was hit by a high current flow through the ac wall power outlet into the ac power adapter, thus damaging its main pcb.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7978047
MDR Text Key124134261
Report Number2017865-2018-15497
Device Sequence Number1
Product Code DRG
UDI-Device Identifier05414734504799
UDI-Public05414734504799
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 12/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberEX1150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/10/2018
Initial Date FDA Received10/18/2018
Supplement Dates Manufacturer Received10/25/2018
12/12/2018
Supplement Dates FDA Received10/26/2018
12/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/06/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-