MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC ACE 36CM W ERG HANDLE; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC ACE 36CM W ERG HANDLE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number ACE36E
Device Problem Failure to Cut (2587)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/26/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # p93789.Investigation summary: the device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched, and with evidence of contact with metal in or out of the operative field.This blade tip portion may have broken off the device during transport to our analysis site.During functional testing on gen11 an alert screen was displayed.A probable cause for the device to stop activating and the gen11 to display an alert screen is blade damage.Due to the device returned condition the tissue effect issue could not be confirmed.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damage blade can result in a broken blade.The batch history record was reviewed and there were no defects, protocols or ncr(s) found during the manufacturing process related to this complaint.
 
Event Description
It was reported that the scalpel was hard to cut during the procedure.Changed to another one to complete the procedure.There was no report on patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC ACE 36CM W ERG HANDLE
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7979228
MDR Text Key124673655
Report Number3005075853-2018-13701
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10705036000938
UDI-Public10705036000938
Combination Product (y/n)N
PMA/PMN Number
K042777
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial
Report Date 10/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Catalogue NumberACE36E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2018
Initial Date Manufacturer Received 10/02/2018
Initial Date FDA Received10/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/11/2017
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
-
-