Model Number 06640613001
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem Cancer (3262)
Event Date 08/15/2018
Event Type  Injury  
Manufacturer Narrative
Per package insert: intended use: anti-thyroid transcription factor-1 (sp141) rabbit monoclonal primary antibody (anti-ttf-1 (sp141) is directed against the thyroid transcription factor-1 (ttf-1) protein.This antibody exhibits a nuclear staining pattern and may be used to aid in the classification of neoplasms of the lung and thyroid.This antibody is intended for qualitative staining in sections of formalin-fixed, paraffin-embedded tissue.This product should be interpreted by a qualified pathologist in conjunction with histological examination, relevant clinical information, and proper controls.This antibody is intended for in vitro diagnostic (ivd) use.Summary and evaluation: expression of ttf-1 has now been identified in tissues and tumors previously thought to be negative with the use of newer clones and detection systems.Literature published over the past few years has demonstrated expression of ttf-1 being detected in neoplastic tissue beyond lung and thyroid.Rationale: assay is a adjunct assay used to aid in the classification of neoplasms of the lung and thyroid.Per medical opinion: case details do not provide evaluation of the h&e, a panel of stains is performed to rule in and rule out various diagnostic potentials.Assay was positive during initial run and re-run supporting no malfunction of assay.Assay ran as intended per package insert.
 
Event Description
Patient went to the hospital.A lung tumor was expected.A lung biopsy was taken.This biopsy was positive for ttf-1 sp141 (lot# (b)(4)) which means an adenocarcinoma of the lung.This sample was very positive, lung control tissue on the same slide stained appropriate as well.This patient was sent to another hospital since tumor outgrowth in the "torax" was diagnosed.A surgery was performed based on the ttf-1 results.Just as control, a ttf-1 staining has been performed after the surgery in this hospital where the surgery has been performed (dako omnis staining platform, ttf-1 dako antibody): staining negative for ttf-1.This means that it could not be an adenocarcinoma of the lung and that the wrong treatment might have been performed.Additional stainings have been performed.This tumor stains positive for b-cell lymphoma on the dako omnis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TTF-1 (SP141) PAB
Type of Device
TTF1
Manufacturer (Section D)
VENTANA MEDICAL SYSTEMS INC
1910 e innovation park dr
tucson AZ 85755 1962
Manufacturer Contact
tim giblin
1910 east innovation park dr
na
tucson, AZ 85755
5208777035
MDR Report Key7982260
MDR Text Key124301205
Report Number2028492-2018-00012
Device Sequence Number1
Product Code NJT
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
CLASS1EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 10/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number06640613001
Device Catalogue Number790-4756
Device Lot NumberE03354
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/25/2018
Initial Date FDA Received10/18/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
-
-