• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH CONFIDENCE SPINAL CMT SYS, 11C; CEMENT, BONE, VERTEBROPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH CONFIDENCE SPINAL CMT SYS, 11C; CEMENT, BONE, VERTEBROPLASTY Back to Search Results
Model Number 283910000
Device Problem No Flow (2991)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that, after 45 seconds of stirring time, the bone cement cannot be extruded.Patient consequence? : no.
 
Manufacturer Narrative
Product complaint #: (b)(4).Udi: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The cement mixer and reservoir were returned, including a limited amount of cement.The returned cement is mostly homogenous with some loose powder.Less than 2 ml of the cement appears to have reached the reservoir.Little direct investigation can be performed as the cement has already solidified.There were no other problems found with the remaining components of the kit.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.The root cause of the cement setting prematurely cannot be determined from the sample and the information provided.A potential root cause may be environment factors related to the storage of the cement.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIDENCE SPINAL CMT SYS, 11C
Type of Device
CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key7996860
MDR Text Key125438852
Report Number1526439-2018-51005
Device Sequence Number1
Product Code NDN
UDI-Device Identifier10705034209623
UDI-Public(01)10705034209623
Combination Product (y/n)N
PMA/PMN Number
K060300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup
Report Date 10/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283910000
Device Catalogue Number283910000
Device Lot Number205367
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2018
Initial Date Manufacturer Received 10/09/2018
Initial Date FDA Received10/23/2018
Supplement Dates Manufacturer Received11/07/2018
11/29/2018
Supplement Dates FDA Received11/19/2018
12/05/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-