Catalog Number PSE45A
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/25/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch number
=
> p57941.Investigation summary
=
> the analysis found that one pse45a device was returned with no apparent damage and with the manual override door out of position; the override lever was up which denotes that the knife was manually returned to home position.It should be noted after the manual override system is used the instrument is disabled and cannot be used for any subsequent firings.A gst45t cartridge reload was received partially fired 1/10 and with 1 driver missing making the reload non-functional.The bailout system was reset and then, it was tested for functionality in the straight position with a test cartridge reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples were noted to have the proper b-form shape.It is possible that while loading the reload, the cartridge was pushed farther back than the cartridge alignment stop windows resulting in the knife pushing the one piece sled forward and locking the cartridge.Although no conclusion could be reached on what caused the drivers to be out of position.In addition it should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure that the device meets the required specifications prior to shipment.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.Additional information requested and received: is the current patient status known? - unknown.Was there any patient consequence or change in the post-operative care of the patient as a result of the event? - unknown.
 
Event Description
It was reported that no advance on the second fire after first use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PWRD ECH FLEX 45MM 34CM SHAFT
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7996926
MDR Text Key124968482
Report Number3005075853-2018-13880
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10705036014317
UDI-Public10705036014317
Combination Product (y/n)N
PMA/PMN Number
K110385
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 07/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/13/2020
Device Catalogue NumberPSE45A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2018
Initial Date Manufacturer Received 10/22/2018
Initial Date FDA Received10/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/13/2017
Is the Device Single Use? Yes
Type of Device Usage Initial