Catalog Number 400SMTXSFT0206
Device Problems Mechanical Jam (2983); Physical Resistance/Sticking (4012)
Patient Problem No Patient Involvement (2645)
Event Date 09/28/2018
Event Type  malfunction  
Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
The patient was undergoing a coil embolization procedure in the middle meningeal artery using penumbra smart coils (smart coil).During the procedure, the physician successfully placed two non-penumbra coils using a non-penumbra microcatheter.While attempting to advance a smart coil out of the introducer sheath into the saline, the physician experienced resistance in the middle of the introducer sheath and the coil became stuck.Therefore, it was not used in the procedure.The procedure was completed an angiography.There was no adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Type of Device
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key8001453
MDR Text Key124974744
Report Number3005168196-2018-02126
Device Sequence Number1
Product Code HCG
UDI-Device Identifier00814548016030
UDI-Public00814548016030
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
K160832
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/08/2023
Device Catalogue Number400SMTXSFT0206
Device Lot NumberF82740
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 10/03/2018
Initial Date FDA Received10/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/09/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-