Model Number P3200K000019
Device Problem Self-Activation or Keying (1557)
Patient Problem No Patient Involvement (2645)
Event Date 10/05/2018
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technical support representative performed troubleshooting over the phone with the account to determine the left foot pedal of the bed needed to be replaced.Per the hill-rom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Hilow motor: examine the motor for the presence and tightness of the attachment hardware.Replace as necessary.Fully raise and lower the bed.If the bed does not fully raise and lower, calibrate the bed position sensor.Make sure there is no friction or abnormal noises and no audible overload indication can be heard during the movement.Make sure the bed not down led lights up on the control pendant and goes out when the bed is in low position.Replace the defective motor(s) in the event of a malfunction.Troubleshoot in the event of a doubt.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in december 2016.It is unknown if the facility performed any other preventative maintenance on this bed.The account replaced the left foot pedal of the bed to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom technical support received a report from the account stating the bed was raising on its own.The bed was located in the biomed department at the account.There was no patient/user injury reported.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE FRAME
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
lori daulton
1069 state route 46 east
batesville, IN 47006
8129312280
MDR Report Key8005572
MDR Text Key125149311
Report Number1824206-2018-00410
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP3200K000019
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/05/2018
Initial Date FDA Received10/25/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-