Patient receiving blood transfusion.About 1.5 hours into blood transfusion, treating rn noticed the blood tubing/blood tubing filter chamber was completely full (it hadn't been) and blood appeared to be leaking down the side of the filter chamber.Transfusion was stopped, charge nurse in to assess situation.Could not tell where blood was leaking from, but it was determined that the iv tubing was no longer a closed system (had been compromised) and transfusion was terminated.Appeared to be a blood tubing malfunction.Lot # on blood tubing is lot (10)dr18b10021.Tubing is baxter clearlink system, y-type blood/solution set with standard blood filter.
|