TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP
|
Back to Search Results |
|
Model Number 816571 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/02/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation is in progress, but not yet concluded.Per the user facility's perfusionist, an "underspeed", "pump jam" and "pump error" were all displayed.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the unit displayed multiple error messages.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
During laboratory analysis, the product surveillance technician (pst) observed the pump to function properly when set to optimal occlusion setting.The pump did produce "under-speed" and "pump jam" error messages during extreme occlusion, which was an expected result.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.Per data log analysis, on (b))6) 2018 the large roller pump logs "belt slip jam status = true", "underspeed (head < demand)", and then eventually "motor fault" as reported.Each of these events occur multiple times.This is an indication of an internal hardware issue with the pump.The service repair technician (srt) was unable to duplicate the reported complaint.The pump operated as intended when set to optimal occlusion settings.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|