• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. FOOTPRINT ULTRA PK SUTURE ANCHOR, 5.5; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. FOOTPRINT ULTRA PK SUTURE ANCHOR, 5.5; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 72202902
Device Problems Break (1069); Device Slipped (1584); Detachment of Device or Device Component (2907)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/05/2018
Event Type  malfunction  
Event Description
It was reported that the anchor slipped off four times.Doctor tightened the stay suture as tight a possible by it didn't help, the anchor kept slipping off.The inserter kept bending in the process and the small tip broke off, the anchor got damaged in the process.A backup device was available to complete the procedure with no significant delay or patient injuries reported.
 
Manufacturer Narrative
Visual inspection and functional testing could not be performed because the device in question will not be returned for evaluation.Thus, the complaint could not be verified and a root cause could not be determined with confidence.It is difficult to perform an accurate evaluation of a failure without having the failed product or the pertinent clinical details to consider.As such the complaint is being closed without conclusion.However, if the product is returned in the future the complaint can be reopened and evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOOTPRINT ULTRA PK SUTURE ANCHOR, 5.5
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key8011665
MDR Text Key125309590
Report Number1219602-2018-01452
Device Sequence Number1
Product Code MBI
UDI-Device Identifier03596010650535
UDI-Public03596010650535
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K093897
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/06/2023
Device Catalogue Number72202902
Device Lot Number50745441
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/05/2018
Initial Date FDA Received10/28/2018
Supplement Dates Manufacturer Received11/20/2018
Supplement Dates FDA Received11/21/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/06/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-