• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT PT/INR CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT PT/INR CARTRIDGE Back to Search Results
Catalog Number 03P89-24
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Information (3190)
Event Date 10/09/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event Description
On (b)(6) 2018, abbott point of care was contacted by a customer regarding i-stat pt/inr cartridges that yielded suspected discrepant results on a patient.There was no patient information, or test/collection times at the time of this report.Method: i-stat, result: 5; lab, 8.There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on information available that suggests the product was not performing within the variability of the assay.It was determined that both the i-stat and lab results provided are above the therapeutic range.The investigation is underway.
 
Manufacturer Narrative
Apoc incident #: (b)(4).The investigation was completed on 11/21/2018.A review of the device history record (dhr) confirmed that the cartridge lot passed release specifications.Retained cartridge testing of pt/inr cartridge lot t18153 met the acceptance criteria outlined in appendix 1 of q04.01.003 rev.Ac (product complaint level 2 and level 3 investigation procedure).No deficiency has been identified for pt/inr cartridge lot t18153.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT PT/INR CARTRIDGE
Type of Device
PT/INR CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key8023524
MDR Text Key126553362
Report Number2245578-2018-00270
Device Sequence Number1
Product Code GJS
UDI-Device Identifier10054749000170
UDI-Public10054749000170
Combination Product (y/n)N
PMA/PMN Number
K020355
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 11/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/29/2018
Device Catalogue Number03P89-24
Device Lot NumberT18153
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/09/2018
Initial Date FDA Received10/31/2018
Supplement Dates Manufacturer Received11/21/2018
Supplement Dates FDA Received11/26/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-