• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) GREENLIGHT MOXY FIBER OPTIC; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) GREENLIGHT MOXY FIBER OPTIC; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0010-2400
Device Problems Fracture (1260); Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/03/2018
Event Type  malfunction  
Manufacturer Narrative
The glass cap shows a circumferential fracture on the distal side of fiber/cap fusion zone at the bevel edge.The fiber proximal to fracture can rotate independently of outer flow tubing.The glass cap exhibits severe devitrification at output window.The metal cap exhibits mild detritus adhesion on surface.The outer flow tubing open end exhibits minor scratch marks.Based on device analysis, the potential for forward firing may exist.The probable cause of the failure is operational context.The cause was determined based on heat accumulation.Cap wear was accelerated due to anatomical/procedural factors (tissue contact and technique) encountered during the procedure which would limit the performance of the fiber.The identified issues noted above may activate the fiberlife function which would modulate the power showing a pulsing beam or system would be placed into standby mode.An evaluation conclusion code of cause cannot be traced to device was assigned to this investigation.
 
Event Description
It was reported that after 359,375 joules and 6:31 minutes of use the fiber broke.The fiber was replaced with a second fiber and after 125,852 joules and 9:45 minutes of use the fiber was cleaned.It was reported that the fiber broke.The procedure was completed with a third fiber.This report is for the first fiber.
 
Manufacturer Narrative
Relevant history: gland volume 120ml reported.This report is for the same event as manufacturer report: 2937094-2018-01076.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENLIGHT MOXY FIBER OPTIC
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer Contact
alyson harris
3070 orchard drive
san jose, CA 95134
4084563300
MDR Report Key8024142
MDR Text Key125670980
Report Number2937094-2018-01075
Device Sequence Number1
Product Code GEX
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/09/2020
Device Model Number0010-2400
Device Catalogue Number0010-2400
Device Lot Number21587795
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/15/2018
Initial Date FDA Received10/31/2018
Supplement Dates Manufacturer Received10/15/2018
Supplement Dates FDA Received10/31/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/09/2018
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-