• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC CAVITRON PLUS ULTRASONIC SCALER WITH TAP-ON TECHNOLOGY; SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC CAVITRON PLUS ULTRASONIC SCALER WITH TAP-ON TECHNOLOGY; SCALER, ULTRASONIC Back to Search Results
Model Number G136
Device Problems Intermittent Continuity (1121); Restricted Flow rate (1248); Overheating of Device (1437); Infusion or Flow Problem (2964)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
While using a g136 scaler, there was intermittent water flow and the handpiece was getting hot; no injury resulted.
 
Manufacturer Narrative
Multiple unsuccessful attempts were made to obtain the device for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAVITRON PLUS ULTRASONIC SCALER WITH TAP-ON TECHNOLOGY
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
MDR Report Key8024752
MDR Text Key125774920
Report Number2424472-2018-00164
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
PMA/PMN Number
K052334
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 01/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG136
Device Catalogue NumberG136
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/01/2018
Initial Date FDA Received10/31/2018
Supplement Dates Manufacturer Received12/11/2018
Supplement Dates FDA Received01/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-