MERIT MEDICAL SINGAPORE PTE LTD. SAFEDRAW BLOOD SUPPLY SYSTEM Back to Search Results
Catalog Number 688037/JPA
Device Problems Fracture (1260); Material Separation (1562)
Patient Problem Blood Loss (2597)
Event Date 10/10/2018
Event Type  malfunction  
Manufacturer Narrative
The suspect device has returned for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event Description
The account alleges that while returning aspirated blood back into the patient after collecting a blood gas sample, a leak occured within the kit and sprayed the nurse with blood.
 
Manufacturer Narrative
The suspect device was returned for evaluation.The device was examined visually.The complaint is confirmed.The root cause is attributed to the manufacturing process.A review of the device history record was performed and no exception documents were found.A review of the complaint database was performed and no similar complaints for this lot number were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFEDRAW BLOOD SUPPLY SYSTEM
Type of Device
BLOOD SUPPLY SYSTEM
Manufacturer (Section D)
MERIT MEDICAL SINGAPORE PTE LTD.
198 yishun ave 7
singapore, 76892 6
SN  768926
Manufacturer (Section G)
MERIT MEDICAL SINGAPORE PTE LTD.
198 yishun ave 7
singapore, 76892 6
SN   768926
Manufacturer Contact
katie swenson cqe,cba,cqpa.
1600 merit parkway
south jordan, UT 84095
8012531600
MDR Report Key8028843
MDR Text Key125922868
Report Number8020616-2018-00066
Device Sequence Number1
Product Code DSK
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K885235
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2020
Device Catalogue Number688037/JPA
Device Lot NumberC1356269
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/17/2018
Initial Date FDA Received11/01/2018
Supplement Dates Manufacturer Received12/03/2018
Supplement Dates FDA Received12/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/25/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-