Additional comments regarding: the lot number could not be specified by the initial reporter.Upon review of this facility's order history, we confirmed the occurrence involves one of the three following lot numbers: w3745130, manufacture date: 07/06/2016, expiration date 07/06/2019.W3782986, manufacture date: 10/11/2016, expiration date 10/11/2019.W3853209, manufacture date: 04/13/2017, expiration date 04/13/2020.Concomitant medical products: boston scientific jagwire wire guide.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The nonconformances documented for the possible lot numbers said to be involved were reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: the device was used outside of stated intended use (abnormal use).The report indicates the slh-1 was used with a wire guide (boston scientific jagwire) to remove a gastric band.The instructions for use states the following intended use: "this device is used to mechanically crush stones in the biliary duct when other methods of endoscopic removal have failed." the instructions for use further states: "do not use this device for any purpose other than the stated intended use." prior to distribution, all soehendra lithotriptor handles are subjected to a visual inspection and functional testing to ensure device integrity.Corrective action: a review of the complaint history was conducted and this represents an isolated occurrence.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.Additional comments regarding this report: based on the information provided that the device was used off-label, a cook representative has been directed to contact the medical facility involved in an effort to promote further education and understanding related to appropriate usage of this product.
|
During an endoscopic removal of a gastric band, the physician used a cook soehendra lithotriptor handle.When they went to crank it, the handle snapped off.This occurred before any pressure was encountered.It was not forceful.They successfully completed the procedure by cranking the device without a handle.A wire guide was being used instead of a basket.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|