Results: dried blood was found on the exterior of the returned indigo system aspiration catheter 6 (cat6) and inside the hub.The cat6 was kinked approximately 1.0 cm from the hub.Conclusions: evaluation of the returned cat6 confirmed that the cat6 was damaged on its proximal end.If the device is forcefully manipulated at extreme angles during removal from the hoop, the device may become kinked.During functional testing, air leaked from the kink near the hub when flushed through the catheter.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
During preparation for a thrombectomy procedure, the hospital technician flushed the indigo system aspiration catheter 6 (cat6) and fluid was noticed to be leaking near the hub of the cat6.Upon inspection, a crack in the catheter was found.The damage to the cat6 was found prior to use.Therefore, the cat6 was not used in the procedure.The procedure was completed using a new cat6.
|