Model Number 57062
Device Problems Break (1069); Crack (1135); Misconnection (1399)
Patient Problem No Information (3190)
Event Date 11/01/2018
Event Type  malfunction  
Manufacturer Narrative
The customer indicated that she disposed of the power supply.A replacement power supply was sent to the customer.This issue with a damaged rev m power supply for the pump in style device was addressed in investigation (b)(4).The investigation found that they were being damaged during shipment from the manufacturer to medela.This damage was causing the plastic housing to fail prematurely when subjected to normal use and foreseeable misuse.The packaging used by the manufacturer to ship the power supply to medela was not robust enough to handle all of the potential shipping, handling, and abuse conditions that could arise from logistics of the consolidation process.As a result of the investigation, the shipping and consolidation process was modified to reduce the handling and potential for double stacking of the skids.The shipping packaging strength was also increased to further protect the power supply during shipping.Since then, as a part of routine continuous improvement activities, a rev p power supply has been distributed to market, manufactured under a revised design and by a different manufacturer.
 
Event Description
On (b)(6) 2018, the customer reported to medela llc that one corner of the power supply housing for her pump in style breast pump was broken, another corner was cracked and the cover would not screw back together.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP IN STYLE ADVANCED BREASTPUMP - BACKPACK
Type of Device
PUMP, BREAST, POWERED
Manufacturer (Section D)
MEDELA LLC
1101 corporate dr
mchenry IL 60050
Manufacturer (Section G)
MEDELA LLC
1101 corporate drive
mchenry IL 60050
Manufacturer Contact
jan kloiber
1101 corporate drive
mchenry, IL 60050
8004358316
MDR Report Key8037285
MDR Text Key128211464
Report Number1419937-2018-00245
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031614
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number57062
Device Catalogue Number57062
Device Lot Number9207010 REV M 3512
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/01/2018
Initial Date FDA Received11/05/2018
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-