Model Number P612 |
Device Problem
Non Reproducible Results (4029)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/19/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event occurred in (b)(6).
|
|
Event Description
|
The customer stated that a screw was loose on the pipettor of the cobas p 612 pre-analytical system.This caused the pipetting system to leak, in turn causing an unspecified number of sample aliquots to be diluted with water.The customer noticed the diluted aliquots and were aware that any results measured from the aliquots must be incorrect.The customer provided an example of one patient sample which had an erroneous result for potassium.The potassium value measured from an affected aliquot was 50 % different than the value measured from the primary tube.No adverse events were alleged.The potassium electrode lot number and expiration date were asked for, but not provided.There was a leakage in the dilutor valve.The field service engineer adjusted the dilutor valve and the issue did not recur.
|
|
Manufacturer Narrative
|
The customer provided another example of a patient sample affected by the issue.Refer to the attachment.The diluted aliquot tube of the sample was tested on an abl90 radiometer analyzer and values from this tube can be seen on the left side.The primary tube of the sample was also tested on the abl90 radiometer analyzer and values from this tube can be seen on the right side.As part of daily operation, the user is required to check the system for possible leakage prior to running the system.Due to missing this check, the leakage was not noticed.A warning label on the device also emphasizes the importance of checking the fluid system for leakage.
|
|
Search Alerts/Recalls
|