Model Number XOM UNK BUR
Device Problems Adverse Event Without Identified Device or Use Problem (2993); No Apparent Adverse Event (3189)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/17/2018
Event Type  malfunction  
Manufacturer Narrative
Analysis found that there was a broken bur inside the nose cone.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The device was returned with no allegation of malfunction.There was no patient impact.
 
Manufacturer Narrative
Analysis found that bur was broken into at least three pieces, the distal end was not returned.The two pieces measured 3/8¿ and 1-3/8¿.The break points were consistent with being bent until broken and there were tool marks along the length of the two pieces.There was excess wear to the tang.With no damage to the nose cone of the hand piece.If information is provided in the future, a supplemental report will be issued.
 
Event Description
On follow up, the manufacturer representative reported that event occurred prior to use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® BUR - UNKNOWN
Type of Device
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
christy cain
6743 southpoint drive north
jacksonville, FL 32216
9043328353
MDR Report Key8045887
MDR Text Key126531012
Report Number1045254-2018-00569
Device Sequence Number1
Product Code EQJ
Combination Product (y/n)N
Reporter Country CodeSP
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberXOM UNK BUR
Device Catalogue NumberXOM UNK BUR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/10/2018
Initial Date FDA Received11/07/2018
Supplement Dates Manufacturer Received03/14/2019
Supplement Dates FDA Received04/11/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-