Lot Number 27216700-334498
Device Problem High Test Results (2457)
Patient Problem Test Result (2695)
Event Date 10/25/2018
Event Type  malfunction  
Event Description
Notified on (b)(6) 2018 by roche diagnostics of "urgent medical device correction", instructed for coaguchek xs pt test strip (b)(4) through (b)(4) if inr >4.5 to obtain venipuncture.Nursing visit on (b)(6) 2018 to obtain inr via coaguchek xs pt test strip with range as above.Inr was 7.4, physician notified, venipuncture obtained, inr: 4.8.Diagnosis or reason for use: coumadin therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS PT TEST STRIPS
Type of Device
TEST, TIME, PROTHROMBIN
Manufacturer (Section D)
ROCHE DIAGNOSTICS GMBH
indianapolis IN 46250
MDR Report Key8046438
MDR Text Key126963960
Report NumberMW5081131
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Nurse
Type of Report Initial
Report Date 11/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Lot Number27216700-334498
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/06/2018
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age65 YR
Patient Weight67
-
-