A product deficiency was identified due to an increase in complaints for failures of the r100 component of the main electronics board (parts 8-35003665-01 and 8-206623-01).The supplier, (b)(4) determined that the issue is from the r100 component of the hettich centrifuge board which was too small to withstand a current spike.Hettich released an enhanced main board revision 06 with a different r100 resistor.It is considered acceptable to continue using the automation system because the issue does not impact operator nor patient safety.The accelerator aps operations manual and the centrifuge operations manual were reviewed and show that troubleshooting and safety information is provided.There have been no reported injuries for this issue.If the electrical components of the main electronics board fail, it will shut down the centrifuge module.If the centrifuge shuts down, the operator will receive an error message.The frequency of this hazard was assessed and it was determined that though there was a trend in complaints, the overall frequency of low had not changed from the predicted frequency as outlined in the risk management file.
|
The customer reported error code 28 [total timeout expired and lid not locked] on the aps centrifuge.The customer cycled power and then was unable to get the centrifuge to turn back on.At that time they noticed an odd smell.Abbott field service measured the voltage and determined a new board was needed.There was no visible charring on the board.There was no impact to patient management, user safety, or facility damage reported.
|