• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9500-45
Device Problem Retraction Problem (1536)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/25/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2018, that on (b)(6) 2018, a needle retraction issue occurred.The sensor was inserted into the abdomen on (b)(6) 2018.No product was provided for evaluation.Confirmation of the problem and probable cause could not be determined.No additional event or patient information is available.
 
Manufacturer Narrative
(b)(4).
 
Event Description
A sensor was returned for evaluation.The device was visually inspected and found that the applicator had a bent pushrod, not allowing the applicator to quickly retract.The reported event of a needle retraction issue was confirmed.The probable cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key8050979
MDR Text Key126665356
Report Number3004753838-2018-137795
Device Sequence Number1
Product Code MDS
UDI-Device Identifier00386270000255
UDI-Public00386270000255
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P170031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number9500-45
Device Catalogue NumberSTS-OR-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/25/2018
Initial Date FDA Received11/08/2018
Supplement Dates Manufacturer Received01/24/2019
Supplement Dates FDA Received01/25/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age37 YR
Patient Weight54
-
-