• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA USA, INC. PROGRAMMING SOFTWARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA, INC. PROGRAMMING SOFTWARE Back to Search Results
Model Number MODEL 3000
Device Problems Computer Software Problem (1112); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/24/2018
Event Type  malfunction  
Manufacturer Narrative
D.Suspect device software version: version 1.5.2.1.
 
Event Description
It was reported that high impedance was detected on a patient's model 102 generator with the m3000 version 1.5.2.1 programming system.It was previously determined through in-house testing that, for model 102/102r generators programmed to output currents > 1ma, system diagnostics using m3000 v1.0 and m250 software would display ok lead impedance while system diagnostics using m3000 v1.5.2.1 software would display a false high impedance.The cause of these false high impedance messages was a software error on m3000 v1.5.1.2 software; when system diagnostics were performed by the user with m3000 v1.5.2.1 software on m102/102r generators (normal mode output current >0 ma), normal mode diagnostics would actually be performed instead.The execution of normal diagnostics instead of system diagnostics is not apparent to the user, and the returned dcdc code was being compared against system diagnostic thresholds.It was determined that m102/102r system diagnostics functioned as expected when the generator was programmed to 0 ma and provided a true impedance value.When the patient's generator was interrogated with a programming system not updated with the v1.5.2.1 software, the patient's lead impedance was within normal limits.Therefore, the high impedance originally observed was determined to be a false result.No further relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key8061531
MDR Text Key127357266
Report Number1644487-2018-02033
Device Sequence Number1
Product Code MUZ
UDI-Device Identifier05425025750429
UDI-Public05425025750429
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 3000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 10/24/2018
Initial Date FDA Received11/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/21/2018
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
Patient Age50 YR
-
-