• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OBERDORF SYNTHES PRODUKTIONS GMBH 2.7MM/3.5MM TI LCP LAT DISTAL FIBULA PLATE 7H/LEFT/125MM; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OBERDORF SYNTHES PRODUKTIONS GMBH 2.7MM/3.5MM TI LCP LAT DISTAL FIBULA PLATE 7H/LEFT/125MM; PLATE, FIXATION, BONE Back to Search Results
Catalog Number 04.112.145
Device Problems Nonstandard Device (1420); Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Additional pro code: hwc.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a synthes employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.A device history record review has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from (b)(6) reports an event as follows: it was reported that during an unknown procedure on an unknown date, a titanium (ti) locking compression plate (lcp) distal fibula plate was noted to have no screw thread.The plate was changed to another one to complete the procedure.There was no adverse consequence to the patient.This report is for a 2.7mm/3.5mm ti lcp lateral distal fibula plate.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h3, h6: part: 04.112.145; lot: h141281; manufacture date: july 15, 2016; manufacture location: elmira synthes usa; expiration date: n/a; noted nonconformances: n/a.A review of the device history record revealed no complaint related anomalies.The device history record shows this lot of 2.7mm/3.5mm ti lcp lat distal fibula product was processed through the normal manufacturing and inspection operations with no rework nor nonconformities noted.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.A review of the raw material device history record revealed this lot met all specifications with no nonconformance noted.This raw material lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.Product will not be received for investigation, but three photos were provided.Investigation site: elmira synthes usa.The visual inspection performed on the provided photos confirms the complaint condition of missing threads as there is no visual presence of threads in the dlh holes (dlh1-dlh5).Relevant actions have been taken to address the issue.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7MM/3.5MM TI LCP LAT DISTAL FIBULA PLATE 7H/LEFT/125MM
Type of Device
PLATE, FIXATION, BONE
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key8061719
MDR Text Key127217633
Report Number8030965-2018-58118
Device Sequence Number1
Product Code HRS
UDI-Device Identifier07611819382855
UDI-Public(01)07611819382855
Combination Product (y/n)N
PMA/PMN Number
K083213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04.112.145
Device Lot NumberH141281
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/26/2018
Initial Date FDA Received11/12/2018
Supplement Dates Manufacturer Received01/16/2019
Supplement Dates FDA Received01/16/2019
Patient Sequence Number1
-
-