Blank fields on this form indicate the information is unchanged, unknown or unavailable.Concomitant medical product: received on: (b)(6) 2018.Investigation: evaluation.A review of the complaint history, device history record, drawing, instructions for use (ifu), manufacturing instructions, quality control, and specifications of the device, as well as a visual inspection, dimensional verification, and functional test, were conducted during the investigation.One catheter was returned in a used, undamaged condition.The visual inspection of the returned device found the suture secure in the threads of the cap.The tubing and cap were also both secure.The dimensions of the device were measure within specifications, so there is no indication from inspection that there are nonconforming devices in the field.The functional test of the device did not confirm the presence of a leak.Additionally, a document based investigation evaluation was performed.The proper procedures are in place to identify and prevent this failure mode prior to device distribution.The risk specifications covering mac-loc drainage catheters include both hub separation and leakage as a potential failure mode.The identified risk controls include manufacturing quality control checks and process validation.The technical files covering mac-loc drainage catheters indicate that the risks associated with these devices are acceptable when weighed against the benefits.A review of the device history record found no related non-conformances with this lot.It should be noted that a review of complaint history found no other related complaints for this lot.There is no evidence of nonconforming devices in house or in field.The instructions for use (ifu) advise the user to inspect the device prior to use to ensure that no damage has occurred.Based on the information provided, inspection of the returned product and the results of our investigation, a definitive root cause could not be established.Appropriate measures have been taken to address this failure mode.The appropriate personnel have been notified.Per the quality engineering risk assessment no further action is required.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|