• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. MEDTRONIC LEAD; PULSE-GENERATOR, PACEMAKER, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. MEDTRONIC LEAD; PULSE-GENERATOR, PACEMAKER, IMPLANTABLE Back to Search Results
Model Number MDT-LEAD
Device Problems Over-Sensing (1438); Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/19/2018
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the right ventricular (rv) lead was not well connected with the implantable pulse generator (ipg).It was also reported there was noise on the rv channel.The rv lead was successfully reconnected with the ipg.The ipg and rv lead remain in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received indicated at follow up check, during a holter testing there were non conducted right atrial (ra) lead p-waves.The physician performed a follow up check but all parameters seems stable.Only amplifying the atrial signal there were isolate double p-wave counting.The sensibility was re-programmed and the double counting disappeared.The ra lead remains in use.
 
Manufacturer Narrative
The lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory was performed and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC LEAD
Type of Device
PULSE-GENERATOR, PACEMAKER, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
lisa robertson
8200 coral sea st ne
mounds view, MN 55112
7635262723
MDR Report Key8066770
MDR Text Key127051548
Report Number2182208-2018-02069
Device Sequence Number1
Product Code DXY
Combination Product (y/n)N
Reporter Country CodeIT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMDT-LEAD
Device Catalogue NumberMDT-LEAD
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/19/2018
Initial Date FDA Received11/13/2018
Supplement Dates Manufacturer Received12/18/2018
12/18/2018
Supplement Dates FDA Received01/14/2019
01/14/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-