Catalog Number 07K70-30
Device Problem Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/06/2018
Event Type  malfunction  
Manufacturer Narrative
Evaluation of the customer issue included a complaint text review, a search for similar complaints, device history review, instrument log review, labeling review, and field data review.The product was not returned.No adverse trend was identified for the customer issue.Device history review did not identify any issues that may have caused the customer issue.Labeling was reviewed and found to be adequate.Log review did not identify any issues that contributed to the customer issue.Abbott link world wide data was used to evaluate historical product field performance.Patient data was analyzed and evaluation indicated that the patient median result for the complaint lot was within the established control limits and performance was comparable to other lots in the field between february 11, 2018 and august 11, 2018.No unusual reagent lot performance was identified.Based on all available information and abbott diagnostics evaluation, no systemic issue was identified and no product deficiency was found.This report is being filed on an international product, list number 7k70 that has a similar product distributed in the us, list number 6c06.Issue for the same patient using a different complaint lot, 85185fn00, was filed under mdr number 3008344661-2018-00094.
 
Event Description
The customer observed falsely depressed results while using architect total psa (prostate specific antigen) reagents.The following data was provided for the same patient.Sid (b)(6) initial 0.83 ng/ml, repeat 0.80, 0.79, 0.74 ng/ml, repeat in another laboratory 0.71, 0.61, 0.83, 0.88 ng/ml.Sid (b)(6) initial 1.19 ng/ml, not repeated.Sid (b)(6) initial 2.02 ng/ml, not repeated.Sid (b)(6) initial 1.16 ng/ml, not repeated.Sid (b)(6) initial 3.38 ng/ml, not repeated.Sid (b)(6) initial 1.09 ng/ml, not repeated.Sid (b)(6) initial 4.57 ng/ml, not repeated.Sid (b)(6) initial 7.23 ng/ml, not repeated.Sid (b)(6) initial 0.74 ng/ml, not repeated.No impact to patient management was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT TOTAL PSA
Type of Device
TOTAL PSA
Manufacturer (Section D)
ABBOTT IRELAND
diagnostics division
finisklin business park
sligo NA
EI  NA
Manufacturer (Section G)
ABBOTT IRELAND
diagnostics division
finisklin business park
sligo NA
EI   NA
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key8066937
MDR Text Key129191484
Report Number3008344661-2018-00095
Device Sequence Number1
Product Code LTJ
Combination Product (y/n)N
Reporter Country CodeAR
PMA/PMN Number
P910007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/03/2019
Device Catalogue Number07K70-30
Device Lot Number84050FN00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/18/2018
Initial Date FDA Received11/13/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/26/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age41 YR
-
-