Catalog Number 8733-7183
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/16/2018
Event Type  malfunction  
Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
 
Event Description
It was reported that the jaw broke off a rongeur during surgery.The procedure was completed using an alternative rongeur without reported patient impacts.
 
Manufacturer Narrative
Udi: na.Additional information: the returned rongeur was examined.The top slide has fractured from the device, making it no longer functional.The cause of this event can likely be attributed to accumulated damage from large applied forces during use over time.A review of the dhr did not find any issues which would have contributed to this event.
 
Event Description
It was reported that the jaw broke off a rongeur during surgery.The procedure was completed using an alternative rongeur without reported patient impacts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RONGEUR PITUITARY CER BLACK SERRATED 3MM
Type of Device
RONGEUR
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
ashley mcpherson
10225 westmoor dr.
na
westminster, CO 80021
3034437500
MDR Report Key8067052
MDR Text Key127214425
Report Number3012447612-2018-00916
Device Sequence Number1
Product Code HTX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8733-7183
Device Lot NumberIM00176
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/16/2018
Initial Date FDA Received11/13/2018
Supplement Dates Manufacturer Received03/06/2019
Supplement Dates FDA Received03/13/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/26/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-