TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP
|
Back to Search Results |
|
Model Number 801041 |
Device Problem
Erratic or Intermittent Display (1182)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/25/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The srt observed the display to not turn on when the pump was plugged in.After tapping on the screen, the display operated normally.
|
|
Event Description
|
The service repair technician (srt) reported that during routine testing of the device at the service center, the pump had intermittent display operation.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.The service repair technician (srt) will replace the display assembly.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|