TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 801188 |
Device Problem
Failure to Calibrate (2440)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/26/2018 |
Event Type
malfunction
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the electronic gas patient system (epgs) had a gas system failure.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Manufacturer Narrative
|
Per the user facility's perfusionist, they attempted to calibrate the electronic patient gas system (epgs) as usual but was unable to after several attempts.The gas lines were checked for leaks, disconnected and reconnected but calibration still would not pass.The epgs was connected to gas lines in another operating room (or), calibration was initiated and passed.It was left running and after some time it was alarming and a red message appeared saying that the gas flow needed to be controlled by the manual controls.She tried to adjust the gas flow manually but no gas was coming from the epgs.Gas lines were checked for leaks, disconnected and reconnected but was unable to resolve the issue.Their circuits have oxygen (o2) and room air going directly from the or supply to the back of the pump.Tubing runs from the back of the pump to the vaporizer, then down to a dwyer air flow meter.From the flow meter the tubing connects to the oxygenator.
|
|
Manufacturer Narrative
|
During laboratory analysis, the product surveillance technician (pst) observed that the gas system failure was a result of a defective internal flowmeter.He connected the electronic patient gas system (epgs) to a system 1 simulator and central control monitor (ccm) with zero air and oxygen pressure.The ccm displayed >10 even though there was no air flow.He replaced the internal flowmeter with a lab use flowmeter and the ccm, with no flow then showed 0 instead of >10.After the 15 minute warmup period, calibration of the epgs was initiated and passed.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Manufacturer Narrative
|
The field service representative (fsr) verified the reported complaint.He replaced the electronic patient gas system (epgs).The unit operated to the manufacturer's specifications.Per the suppliers evaluation, the unit was received reading 12.80 standards liters per minute (slpm) at zero flow.Analog outputs track correctly.Absolute pressure reading 17.95 at atmosphere (13.64 pounds per square inch (psia).Register inspection reveals differential output has shifted since time of production (sensor signal counts shifted from 9874 to 65362 and accounts for erroneous flow readings).Sensor bridge values have altered since production at factory (39100 to 32755) and accounts for erroneous pressure reading.These readings point to failure/detect in differential pressure sensor.Unfortunately, the sensor was damaged when the casing was removed so full analysis of the sensor cannot be conducted, nevertheless, a loose wire bond was identified by manual pull test.
|
|
Search Alerts/Recalls
|
|
|