• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801188
Device Problem Calibration Problem (2890)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Per the manufacturer's technical support specialist, the user facility's perfusionist inadvertently changed the setting from ¿air¿ to ¿95-5¿ causing a calibration failure message and the data logs to show ¿o2 sensor and blender disagree¿ message.The setting was changed back to ¿air¿, performed calibration and the epgs worked as intended.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the electronic patient gas system (epgs) lost its calibration.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key8084568
MDR Text Key128056312
Report Number1828100-2018-00596
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799000588
UDI-Public(01)00886799000588(11)180309
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801188
Device Catalogue Number801188
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/29/2018
Initial Date FDA Received11/19/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/09/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-