• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE TORQUE DEFINING SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE TORQUE DEFINING SCREW Back to Search Results
Catalog Number 300-11-00
Device Problem Material Separation (1562)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/08/2018
Event Type  Injury  
Manufacturer Narrative
Pending evaluation.
 
Event Description
Index surgery: (b)(6) 2018.Glenoid disassociated from the cage and one peripheral peg.The glenoid poly component and two peripheral pegs is positioned posterior to the humerus/joint and the cage and one peripheral peg is still stuck in the bone of the glenoid.The humeral head was dis-impacted and the stem came out with it.The primary glenoid component was removed from the joint and the cage and peripheral peg was removed from the glenoid bone.The glenoid was then reamed and re-prepped for re-insertions of an all poly glenoid.A medium beta all poly glenoid implant was cemented in.The humerus was broached up to a 8mm stem and a 0 replicator plate was used.Then a 44mmx17mm humeral head was used for final implants.
 
Manufacturer Narrative
The engineering evaluation noted that the revision reported was stated to be the result of disassociation of the center cage and one of the three peripheral pegs from the polyethylene body of the glenoid component; the result of this analysis confirmed that observation.The cause of this disassociation was inconclusive though it was likely cause by damage to the central peg at the time of the glenoid implantation or by eccentric humeral head edge loading, as evidence by the burnishing along the peripheral of the glenoid articular body.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE TORQUE DEFINING SCREW
Type of Device
TORQUE DEFINING SCREW
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66th court
gainesville FL 32653
MDR Report Key8084838
MDR Text Key127593844
Report Number1038671-2018-00876
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 03/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number300-11-00
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/08/2018
Initial Date FDA Received11/19/2018
Supplement Dates Manufacturer Received11/08/2018
Supplement Dates FDA Received03/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age48 YR
Patient Weight116
-
-