• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER Back to Search Results
Model Number EX1150
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/31/2018
Event Type  malfunction  
Manufacturer Narrative
Final analysis found burn marks on the printed circuit board (pcb) and the ac power adapter which contributed to the field complaint of power up anomaly.Abbott conducted a review of complaint files under a non-product capa and identified that a report was required to be filed as a result of the subject event.Distributed devices are not affected by the capa.Additionally, no new risks were identified, and no product-related actions are required as a result of this report submission.Furthermore, product performance trending is not impacted by this report, and no new mitigation actions are required as a result of this report.
 
Event Description
It was reported that the patient's transmitter would not power up.The transmitter's power plugs were blackened from a storm.A new transmitter was ordered.There were no patient consequences.
 
Manufacturer Narrative
Correction: manufacturing date added to report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8093157
MDR Text Key128022401
Report Number2017865-2018-17492
Device Sequence Number1
Product Code DRG
UDI-Device Identifier05414734504799
UDI-Public05414734504799
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEX1150
Device Catalogue NumberEX1150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/31/2018
Initial Date FDA Received11/20/2018
Supplement Dates Manufacturer Received01/21/2019
Supplement Dates FDA Received01/21/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/10/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-