Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
Visual examination confirmed the upper tang/prong of the threaded inserter has broken off; the broken piece is missing and not returned for analysis.Fracture initiation appears to be located at the base of the tang, consistent with bend stress overload.Optical examination of the fracture surface reveals a fairly quasi-brittle fracture, with no indication of fatigue, and a shear lip which suggests the direction of fracture.The above observations are consistent with bend stress overload.If information is provided in the future, a supplemental report will be issued.
|