Catalog Number 2982001
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/29/2018
Event Type  malfunction  
Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient presented with stenosis; and underwent oblique lumbar interbody fusion.Intra-op, the distal tip of the inserter broke during cage insertion.The product came in contact with the patient.No fragment of the broken instrument remained in the patient.No patient complications were reported as a result of this event.
 
Manufacturer Narrative
Visual examination confirmed the upper tang/prong of the threaded inserter has broken off; the broken piece is missing and not returned for analysis.Fracture initiation appears to be located at the base of the tang, consistent with bend stress overload.Optical examination of the fracture surface reveals a fairly quasi-brittle fracture, with no indication of fatigue, and a shear lip which suggests the direction of fracture.The above observations are consistent with bend stress overload.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPINAL VERTEBRAL BODY REPLACEMENT DEVICE
Type of Device
SPINAL VERTEBRAL BODY REPLACEMENT DEVICE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
stacie ziemba
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key8099748
MDR Text Key128197811
Report Number1030489-2018-01527
Device Sequence Number1
Product Code MQP
UDI-Device Identifier00885074549736
UDI-Public00885074549736
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062133
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2982001
Device Lot NumberPR11C019
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/29/2018
Initial Date FDA Received11/23/2018
Supplement Dates Manufacturer Received01/23/2019
Supplement Dates FDA Received01/31/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age46 YR
-
-