• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON MB PATELLA PA A32; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON MB PATELLA PA A32; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, C Back to Search Results
Catalog Number 5554L320
Device Problems Noise, Audible (3273); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problems Inflammation (1932); Pain (1994); Swelling (2091); Weakness (2145); Ambulation Difficulties (2544)
Event Date 12/01/2016
Event Type  Injury  
Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.Based on the device identification the complaint databases were reviewed for similar reported events regarding loosening.There have been no other events for the lot referenced.
 
Event Description
As reported through mw5078695 "patient reports she was unable to get up from her bed one morning due to severe pain in her knee.She reports hearing clicks while walking and has experienced weakness, swelling and buckling.Her orthopedic surgeon saw "something sticking out of her bone" and mentioned that "something pulls away" when she moves her knee.The patient is wearing a knee brace for support and can no longer hear the clicking noise; however, her other symptoms persist" update (b)(6) 2018: right knee loosening.
 
Event Description
As reported through mw5078695 "patient reports she was unable to get up from her bed one morning due to severe pain in her knee.She reports hearing clicks while walking and has experienced weakness, swelling and buckling.Her orthopedic surgeon saw "something sticking out of her bone" and mentioned that "something pulls away" when she moves her knee.The patient is wearing a knee brace for support and can no longer hear the clicking noise; however, her other symptoms persist." update 29-oct-2018: right knee loosening.
 
Manufacturer Narrative
An event regarding audible noise involving a triathlon patella was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as no product was returned for evaluation.Medical records received and evaluation: a review of the provided medical records by a clinical consultant indicated: on (b)(6) 2018 a note states, ¿seen (b)(6) 2018 and diagnosed with unstable right tkr.Reports ascending and descending stairs virtually impossible.Reports clicking sound and sensation with weightbearing.Physical examination: 0° to 120° with jog of laxity in full extension and 90°.Ap drawer test negative.Plan: bledsoe brace, strengthening exercises, return in two months.¿ no documented follow-up subsequent to this visit is available.X-ray printouts include multiple views of the right knee, all undated, demonstrating a right posterior stabilized total knee arthroplasty, uncemented, with four screws in the tibial baseplate.The knee is reduced and the components are in nominal position.There is no evidence of loosening or pathology noted.On (b)(6) 2018 a note states, ¿seen (b)(6) 2018 and diagnosed with unstable right tkr.Reports ascending and descending stairs virtually impossible.Reports clicking sound and sensation with weightbearing.Physical examination: 0° to 120° with jog of laxity in full extension and 90°.Ap drawer test negative.Plan: bledsoe brace, strengthening exercises, return in two months.¿ no documented follow-up subsequent to this visit is available.X-ray printouts include multiple views of the right knee, all undated, demonstrating a right posterior stabilized total knee arthroplasty, uncemented, with four screws in the tibial baseplate.The knee is reduced and the components are in nominal position.There is no evidence of loosening or pathology noted.The event description states, ¿she reports having clicks while walking and has weakness, swelling, and buckling wearing a knee brace for support and can no longer hear the clicking.Device history review: indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: a review of the provided medical records by a clinical consultant indicated: a review of the provided medical records by a clinical consultant indicated: on (b)(6) 2018 a note states, ¿seen (b)(6) 2018 and diagnosed with unstable right tkr.Reports ascending and descending stairs virtually impossible.Reports clicking sound and sensation with weightbearing.Physical examination: 0° to 120° with jog of laxity in full extension and 90°.Ap drawer test negative.Plan: bledsoe brace, strengthening exercises, return in two months.¿ no documented follow-up subsequent to this visit is available.X-ray printouts include multiple views of the right knee, all undated, demonstrating a right posterior stabilized total knee arthroplasty, uncemented, with four screws in the tibial baseplate.The knee is reduced and the components are in nominal position.There is no evidence of loosening or pathology noted.On (b)(6) 2018 a note states, ¿seen (b)(6) 2018 and diagnosed with unstable right tkr.Reports ascending and descending stairs virtually impossible.Reports clicking sound and sensation with weightbearing.Physical examination: 0° to 120° with jog of laxity in full extension and 90°.Ap drawer test negative.Plan: bledsoe brace, strengthening exercises, return in two months.¿ no documented follow-up subsequent to this visit is available.X-ray printouts include multiple views of the right knee, all undated, demonstrating a right posterior stabilized total knee arthroplasty, uncemented, with four screws in the tibial baseplate.The knee is reduced and the components are in nominal position.There is no evidence of loosening or pathology noted.The event description states, ¿she reports having clicks while walking and has weakness, swelling, and buckling wearing a knee brace for support and can no longer hear the clicking.It is also noted that the patient inquired if the product was part of the recall.A letter from stryker orthopedics dated october 29, 2018 states, ¿none of these products have been involved in a recall.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON MB PATELLA PA A32
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, C
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8105554
MDR Text Key128397111
Report Number0002249697-2018-03824
Device Sequence Number1
Product Code MBH
UDI-Device Identifier07613327049893
UDI-Public07613327049893
Combination Product (y/n)N
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 01/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2016
Device Catalogue Number5554L320
Device Lot NumberSY4KA
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/29/2018
Initial Date FDA Received11/27/2018
Supplement Dates Manufacturer Received12/19/2018
Supplement Dates FDA Received01/16/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age79 YR
Patient Weight61
-
-