The customer complained of an issue with the test strip on a cobas u 411 analyzer with software version 3.2.0 that has been occurring 1-2 times per week for approximately 1 month.The test strip was pushed into the slot but was not detected by the "strip check" function performed by the analyzer so the customer inserted a 2nd test strip.Since the first test strip was not processed, there was a mismatch between patient results.The specific results were requested but have not been provided.There was no allegation that an adverse event occurred.No instrument alarms or flags were displayed when the test strip was not detected.The field service engineer (fse) visited the customer site and found that the "strip check" function was working fine but on rare occasions the test strip was not detected.There will be no alarm and the order list does not move forward, but the customer will insert a new test strip.The fse advised the customer to verify that the "strip check" occurs prior to inserting a new test strip.The fse noted during random testing that when the test strip was not detected, he noticed the test strip was not close enough to the "strip check" sensor.The fse thinks the issue is with the test strip transfer system and not the sensor.The fse replaced the test strip transfer and lubricated the transport mechanism.Investigations are ongoing.
|