Investigation results: one 8360-10, 5mmx36cm atraumatic dual action grasper device returned.The complaint allegation stated: "the jaws of the device fell off inside the patient." the device components separated during an abdominal procedure.The identification and logo areas shows signs of bead blasting.Visual inspection confirmed broken jaw components.The jaws themselves were absent upon receipt, however, there is tape and a small packet attached to the device that was empty.Regardless, the distal portion of the device has been broken.The tip remaining is twisted and no longer aligned/centered with the luer port.The symptoms indicate excess torque was used.Per instructions for use: "aesculap prestige endoscopic graspers are designed to grasp soft tissue structures during endoscopic procedures.Aesculap endoscopic graspers are designed to be used through a cannula, or port, commonly called a trocar sleeve.Aesculap prestige endoscopic graspers are delicate surgical instruments.Any use of an instrument for a task other than its intended purpose will usually result in a damaged or broken instrument.Lateral pressure on the instrument when inserting or removing it may damage the shaft or the working tip." the condition may also occur during cleaning.
|