The customer stated that the cobas c8100 system is sending sample tubes to their cobas 8000 analyzer with sample volumes that are too low.This issue led to a clogged ise probe on their cobas 8000 analyzer.The probe was clogged with separator gel, which led to the correction of ise indirect na for gen.2 results for 88 patient samples.The customer provided an example of one patient sample which had an erroneous result for na that was reported outside of the laboratory.The sample initially resulted with an na value of 141 mmol/l.The sample was repeated, resulting as 149 mmol/l.The repeat result was believed to be correct.The patient was not adversely affected.The na electrode lot number and expiration date were asked for, but not provided.The field service engineer checked the c8100 system and it was found to be working correctly.The customer informed the engineer that the system was working well as far as he knew.The field application specialist checked the system and stated that sample volumes were differing by volumes of 500 ul to 1 ml and 2 ml.The tubes that were processed on the c8100 were sent to the cobas 8000 analyzer for testing, but none of these tubes had sample short errors or were rejected by the c8100 system.
|
The investigation found the chuck in the automatic centrifuge unit was not properly adjusted, the middleware was not updated to the latest software version, and the laser liquid level detection was not recalibrated after the c8100 software update.Field service replaced and adjusted the chuck assembly.The system ran without errors.
|