• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION CPAP PRO; VENTILATOR, NON-CONTINUOUS (RESPIRATOR), PRODUCT CODE: BZD,,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION CPAP PRO; VENTILATOR, NON-CONTINUOUS (RESPIRATOR), PRODUCT CODE: BZD,, Back to Search Results
Model Number GBX400S15
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problem Death (1802)
Event Date 11/28/2018
Event Type  Death  
Event Description
The manufacturer became aware of a patient death.A family member is alleging that the continuous positive airware pressure (cpap) device may have caused or contributed to the patient's death.The date of death is unknown and there has been no other information provided to date.It is unknown if the patient was wearing the cpap device at the time of death.The investigation is on-going.Upon completion of the investigation, a follow up report will be filed.
 
Manufacturer Narrative
The manufacturer received the devices for investigation.There was no evidence of a failure of the received devices that would have caused or contributed to the reported event.There were no operational issues found.The devices performed as intended and passed all testing.The philips respironics dreamstation systems deliver positive airway pressure therapy for the treatment of obstructive sleep apnea in spontaneously breathing patients weighing over (b)(6).It is for use in the home or hospital/institutional environment.The user is cautioned "contact your health care professional if symptoms of sleep apnea recur.If you notice any unexplained changes in the performance of this device, if it is making unusual or harsh sounds, if it has been dropped or mishandled, or if the enclosure is broken, disconnect the power cord from the therapy device and discontinue use.Contact your home care provider." this device is not intended for life support.Based on the available information, the manufacturer concludes that no further action is necessary at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION CPAP PRO
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR), PRODUCT CODE: BZD,,
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
adam price
312 alvin drive
new kensington, PA 15068
MDR Report Key8120172
MDR Text Key128903703
Report Number2518422-2018-02839
Device Sequence Number1
Product Code BZD
UDI-Device Identifier00606959023482
UDI-Public00606959023482
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberGBX400S15
Device Catalogue NumberGBX400S15
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/31/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/28/2018
Initial Date FDA Received11/30/2018
Supplement Dates Manufacturer Received01/03/2019
Supplement Dates FDA Received01/08/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/09/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Death;
-
-