• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROMODULATION ENTERRA; INTESTINAL STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROMODULATION ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 37800
Device Problems Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Twitching (2172)
Event Date 11/28/2018
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 4351-35, serial# unknown, explanted: (b)(6) 2018, product type: lead.Other relevant device(s) are: product id: 4351-35, serial/lot #: unknown.(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional (hcp) via a manufacturer representative regarding a patient with an implantable neurostimulator (ins).It was reported that the patient experienced twitching of the abdomen during normal use.It was unknown what environmental/external/patient factors may have led or contributed to the issue.Diagnostics/troubleshooting performed included impedance testing and shutting off the device.It was noted that a replacement occurred on (b)(6) 2018, and the leads were cut and the electrodes remained implanted.The issue was resolved at the time of the report.No further complications were reported/anticipated.
 
Manufacturer Narrative
Adverse event or product problem updated.Device code c62973 no longer applies.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information from the manufacture representative (rep) reported the patient had two leads and the battery that were removed and replaced.The cause of the twitching was unknown but resolved.The rep noted the leads were intentionally cut per the physician¿s discretion.There were no further complications that have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
MDR Report Key8122372
MDR Text Key129018362
Report Number3007566237-2018-03473
Device Sequence Number1
Product Code LNQ
Combination Product (y/n)N
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/28/2018
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/28/2018
Initial Date FDA Received12/01/2018
Supplement Dates Manufacturer Received12/31/2018
Supplement Dates FDA Received01/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age56 YR
-
-