• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SHUNT SENSOR SYS500; BLOOD GAS MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SHUNT SENSOR SYS500; BLOOD GAS MONITOR Back to Search Results
Model Number CDI510H
Device Problem Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 11/13/2018
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).Results: results pending completion of investigation.Conclusions: conclusion not yet available.
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during out of box, a leakage was found when the shunt sensor was unpacked.No patient involvement.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on december 2, 2018.  (b)(4).All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.(b)(4).The returned sample was visually inspected and was noted to have dried buffer on the outside of the large bore cap and inside the threads.It was leak tested by connecting with the calibrated manometer, submerged into a water bath, pressurized up to 1030 mmhg, and no leaks were noted.The large blue cap was then loosened and retightened with a calibrated torque wrench to production specification.It was pressurized with air submerged into a water bath, and pressurized up to 1030 mmhg.No leaks were noted.A retention sample from the same lot number was visually inspected and confirmed to have no traces of buffer on the outside of the unit or inside of the pouch.It was also leak tested, and no leaks were noted.The complaint was not confirmed, and a definitive root cause was not able to be determined.It is possible that the buffer solution was not completely wiped off the product after buffer fill during manufacturing.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHUNT SENSOR SYS500
Type of Device
BLOOD GAS MONITOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key8123148
MDR Text Key129057965
Report Number1124841-2018-00319
Device Sequence Number1
Product Code DRY
UDI-Device Identifier00699753160767
UDI-Public(01)00699753160767
Combination Product (y/n)N
PMA/PMN Number
K972962
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 01/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2019
Device Model NumberCDI510H
Device Catalogue NumberN/A
Device Lot NumberWE23G
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/14/2018
Initial Date FDA Received12/02/2018
Supplement Dates Manufacturer Received11/26/2018
01/10/2019
Supplement Dates FDA Received12/11/2018
01/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-