• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS DIVISION CELL-DYN EMERALD; AUTOMATED HEMATOLOGY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS DIVISION CELL-DYN EMERALD; AUTOMATED HEMATOLOGY ANALYZER Back to Search Results
Catalog Number 09H39-01
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
All available patient information was included.Additional patient details are not available.An evaluation is in process.A final report will be submitted when the evaluation is complete.
 
Event Description
The customer observed a falsely elevated platelet result on the cell-dyn emerald analyzer.The following data was provided: initial 1,000,000 k/ul, repeated too high to report, repeated again 415, 300 k/ul.There was no impact to patient management reported.
 
Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs, a review of labeling and instrument service.No adverse trend was identified for the customer's issue.Labeling was reviewed and found to be adequate.The issue was resolved by replacing the o-rings kit and 1l straw.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN EMERALD
Type of Device
AUTOMATED HEMATOLOGY ANALYZER
Manufacturer (Section D)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
MDR Report Key8124442
MDR Text Key129444984
Report Number2919069-2018-00059
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier00380740020088
UDI-Public00380740020088
Combination Product (y/n)N
PMA/PMN Number
K081495
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 01/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number09H39-01
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/09/2018
Initial Date FDA Received12/03/2018
Supplement Dates Manufacturer Received12/30/2018
Supplement Dates FDA Received01/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-