• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOVIVE SAFETY PEG KIT; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOVIVE SAFETY PEG KIT; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number UNK-M-57690
Device Problem Difficult to Advance (2920)
Patient Problem Pain (1994)
Event Date 11/01/2018
Event Type  malfunction  
Manufacturer Narrative
Date of event was approximated to (b)(6) 2018 as no event date was reported.The complainant was unable to provide the suspect device upn and lot number; therefore, the lot expiration and device manufacture dates are unknown at this time.(b)(4).The complainant indicated that the device will not be returned for evaluation as it is implanted in the patient; therefore a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that an endovive safety peg kit pull method was used during a percutaneous endoscopic gastrostomy procedure.The exact procedure date was unknown; however it was reported that it was placed a week before the event happened.According to the complainant, the patient went to the emergency room due to pain at the peg site.Patient also complained of leaking around the tube and the external bolster was tight against the skin.Ct scan was ordered and revealed that there was inflammation of the fat between the stomach and the skin.There was tenting of the gastric mucosa.The boston peg tube external bolster was extremely difficult to loosen, wherein it stretched the peg tube when trying to loosen the external bolster.They used two clamps, one of which was grasping the tube between the skin and the other clamp was used to grasp the external bolster which induced pain to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOVIVE SAFETY PEG KIT
Type of Device
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key8125840
MDR Text Key129084634
Report Number3005099803-2018-61705
Device Sequence Number1
Product Code KNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUNK-M-57690
Device Catalogue Number57690
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/05/2018
Initial Date FDA Received12/03/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-