In anticoagulation clinic we use coaguchek xs point of care machine to monitor and adjust anticoagulation therapy with warfarin for pts with all indications for warfarin therapy.Recently we received a notice (not a recall, just a notice) from roche, the mfr of the test strips used in the machine, that when they re-calibrated the strips to meet new who standards, abnormally high inr results occurred.They are re-calibrating the new lots of test strips back to the old who standard, but have not recalled the erroneous lots.The notice advised confirming inr results >4.5 with an alternate method of testing, however in our clinic we have had pts with point-of-care results of >6 with venipuncture inr results in the 3 range (which is within therapeutic range for many of our pts).We believe the strips may not be accurate even below the roche-designated cut-off of inr of 4.5.Medication administered to or used by the pt: no.(b)(6).Access number: (b)(4).
|