It was reported that the deployment was successful; however, one day after, the anal side of stent was found being not extended.It is hard to review suspected device's dhr, because it is not confirmed serial no.Colon structure where stent was implanted is curvy.Also, based on the description about receipt the deployment fail, which was written that "physician encountered very strong resistance", it seems that it was hard to expand successfully due to pressed by the patient's lesion condition.However, it is hard to identify the exact cause because there is no disease information for the patient and no photo information from the time of the procedure.This suspected device is not registered in the us but we will continuously monitor the same or similar customer complaints through accurate analyses.
|