TAEWOONG MEDICAL CO.,LTD. NITI-S ENTERAL COLONIC UNCOVERED STENT; COLONIC STENT Back to Search Results
Model Number CXDT2208
Device Problem Activation Failure (3270)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/21/2018
Event Type  malfunction  
Manufacturer Narrative
It was reported that the deployment was successful; however, one day after, the anal side of stent was found being not extended.It is hard to review suspected device's dhr, because it is not confirmed serial no.Colon structure where stent was implanted is curvy.Also, based on the description about receipt the deployment fail, which was written that "physician encountered very strong resistance", it seems that it was hard to expand successfully due to pressed by the patient's lesion condition.However, it is hard to identify the exact cause because there is no disease information for the patient and no photo information from the time of the procedure.This suspected device is not registered in the us but we will continuously monitor the same or similar customer complaints through accurate analyses.
 
Event Description
On (b)(6) 2018: cxdt2208 was placed in the patient.On (b)(6) 2018: the anal side of cxdt2208 stent was found being not extended.Another stent (niti-s) was used to finish the procedure but it had stent deployment failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NITI-S ENTERAL COLONIC UNCOVERED STENT
Type of Device
COLONIC STENT
Manufacturer (Section D)
TAEWOONG MEDICAL CO.,LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
KS  10022
Manufacturer (Section G)
TAEWOONG MEDICAL CO.,LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
KS   10022
Manufacturer Contact
park
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
MDR Report Key8132122
MDR Text Key129888549
Report Number3003902943-2018-00032
Device Sequence Number1
Product Code MQR
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCXDT2208
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/19/2018
Initial Date FDA Received12/05/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-