• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-02H
Device Problem Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/14/2018
Event Type  malfunction  
Event Description
The recipient reportedly experienced a potential device failure.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
The external visual inspection revealed a slice in the fantail region prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock could not be obtained at any spacing.The no lock condition prevented some electrical tests from being performed.The device passed some of the electrical tests performed.The device failed the residual gas analysis test.The internal visual inspection noted silver migration between some electrical components.This device had moisture that exceeded the residual gas analysis test limit.The source of the problem was a feedthru hermeticity issue from one feedthru vendor.A corrective action was implemented.Feedthru assemblies from this vendor are no longer used.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
delilah garcia
28515 westinghouse place
valencia, CA 91355
MDR Report Key8137764
MDR Text Key129756261
Report Number3006556115-2018-00592
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/01/2006
Device Model NumberCI-1400-02H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/06/2018
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/10/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/02/2004
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age17 YR
-
-