The reported cable device was not returned to olympus for evaluation.The cause of the reported complaint cannot be confirmed.Despite several followup attempts, there is no further information regarding the procedure, the location of reported burning / melting on the device, the service age of the device, or the handpiece involved.As preventive measures, the cable instructions state, ¿do not use the hf cable after one year of use.¿ ¿visually inspect the cable and the plugs for irregularities on the surface.¿ and ¿do not use a cable with brittle or defective insulation.Replace the cable.¿ the instructions also warn, ¿use only compatible equipment¿ listed in the instructions, and ¿the maximum output voltage of the hf unit must not exceed the lowest maximum rated voltage of any of the hf components used during the procedure.¿ the user facility was also advised of the one year maximum usage recommendation at the time that the event was reported to olympus.
|